As the national metrology institute of Brazil, Inmetro is responsible for the maintenance and support given to clients in Brazil and abroad. In order to implement its services with high quality performance, Inmetro develops systems which will provide the dissemination of the basic magnitudes. Corroborating to this mission, the Voltage and Current Laboratory (Latce) presents its standard for ac-dc current transfer based on Single Junction Thermal Converters (SJTCs), the behaviour of Latce’s standards, and the recent upgrades in the system. Responsible for the calibration of both voltage and current instruments, such as accurate multifunction calibrators and multimeters for other laboratories or industry, the transfer standards are between the first levels of the traceability pyramid for the dissemination of time-dependent voltage and current magnitudes. The system for ac-dc current transfer measurement has been improved since BIPM international key comparison of ac-dc current transfer standards – CCEM K12, by changes in the setup and uncertainty budget reevaluation. The aim of this paper is to show how Latce managed to reduce uncertainty magnitude.

Keywords
current, ac-dc, uncertainty
Download
IMEKO-TC4-2011-060.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2011
Technical Committee
TC4
Email
eventos@metrologia.org.br
Place
Natal, BRAZIL
Time
27 September 2011 - 30 September 2011
Website
http://www.metrologia.org.br/metrologia2011/

Back to the proceedings