Skip to main content

ON THE IMPACT OF ANEMOMETER SIZE ON THE VELOCITY FIELD IN A CLOSED WIND TUNNEL

I. Care, M. Arenas

Abstract

In the present paper, experimental and numerical investigations of the flow around different types and sizes of anemometers are presented and discussed. The measurements of the flow field at different distances upstream to the anemometer are performed with a laser Doppler Anemometer. The computational results are in rather good agreement with the experimental ones and show that anemometers may induce a strong distortion of the velocity field, even far upstream the anemometer.

Keywords
anemometer size, closed wind tunnel, LDA
Download
IMEKO-TC9-2013-033.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2013
Technical Committee
TC9
Place
Paris, FRANCE
Time
24 September 2013 - 26 September 2013
Website
http://www.flomeko2013.fr

Back to the proceedings