Skip to main content

DROP VOLUME ESTIMATION OF INTRAVENOUS SET USING GRAVIMETRIC METHOD

K. Leetang, P. Samanpiboon, T. Chinarak

Abstract

Intravenous Administration Set (IV Set) is one of medical devices that have been generally used in hospitals. Then, the primary standard calibration system appeared in this paper was set according to the IEC60601-2-24. The flow rate, volume and mass of distilled water which flowing through different IV sets are measured by using 1 mg resolution weighing balance, photoelectric sensors and data acquisition system. This is paper presents drop volume estimation compared with three different methods; surface tension calculation, drop counted by photoelectric sensor and gravimetric method. The results show comparison of the drop volume with the three different methods and error less than 5% for each IV set type. The uncertainty of measurement is 0.752% of reading (<i>k</i> = 2).

Keywords
primary standard, calibration, IEC60601-2-24, drop volume estimation, gravimetric
Download
IMEKO-TC9-2013-041.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2013
Technical Committee
TC9
Place
Paris, FRANCE
Time
24 September 2013 - 26 September 2013
Website
http://www.flomeko2013.fr

Back to the proceedings