Skip to main content

DYNAMIC TEMPERATURE EFFECTS IN A CLEARANCE-SEALED PISTON PROVER FOR GAS FLOW MEASUREMENTS

J Kutin, G Bobovnik, I Bajsić

Abstract

This paper deals with the dynamic temperature effects in a high-speed, clearance-sealed realization of a piston prover. A dynamic mathematical model was built to study the temperature variations and to estimate their influence on the flow measurements. The mathematical model is formulated on the basis of lumped-element models of the piston and the gas cavity. The energy balance for the gas cavity includes the convective heat exchange with the surroundings. One of the potential effects results from the temperature differences betwen the inlet gas flow and the cylinder wall. In addition, the piston prover's operation generates dynamic temperature variations which are mainly related to the pressure change due to the flow redirection and to the pressure oscillations due to the piston's resonance effects.

Keywords
dynamic temperature effects, piston prover
Download
IMEKO-TC9-2013-054.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2013
Technical Committee
TC9
Place
Paris, FRANCE
Time
24 September 2013 - 26 September 2013
Website
http://www.flomeko2013.fr