Skip to main content

PRECISE MASS AND VOLUME DETERMINATION OF THE SILICON SINKER FOR SEA WATER DENSIMETRY

Y. Kayukawa, Y. Fujita

Abstract

In order to improve measurement performance in sea water densimetry which is important in global environmental monitoring, a hydrostatic weighing system is under development in the NMIJ. A 100 g silicon sinker is used as a density reference and it is required to calibrate its mass and volume with higher accuracy that the conventional calibration procedures. In the present study, sub multiple mass measurement was conducted by using 1 kg silicon sphere and newly designed silicon disc weight set. The mass uncertainty was decreased to less than 0.01 mg for the silicon sinker. The volume of the sinker was measured by using a pressure-of-floatation method (PFM) to realize the relative uncertainty of the volume being 0.17 ppm.

Keywords
silicon single crystal, density, mass, submultiple, hydrostatic weighing
Download
IMEKO-TC3-2014-022.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
TC3 Conference 2014
Linked User
Dirk Röske
Technical Committee
TC3
Place
Cape Town, SOUTH AFRICA
Time
3 February 2014 - 6 February 2014
Website
http://conferences.imeko.org/index.php/tc3-5-22/2014

Back to the proceedings