A NEW CALIBRATION DEVICE FOR SILK SCREEN TENSION METER
Abstract
In this paper, a calibration device for silk screen tension meter is designed. The device has a measurement range from 5 to 100 N/cm. It can achieve simultaneous loading in <i>x</i> and <i>y</i> directions by servo motors. For loading accuracy, two high precisionforce sensors are utilized in the closed-loop control system for each direction. To verify the device, it is utilized to calibrate a commercial available silk screen tension meter. The results from the device are compared with that from the Swiss Federal Office of Metrology. It demonstrates that the max error is 0.5 %.
Event details
- Event
- TC3 Conference 2014
- Linked User
- Dirk Röske
- Technical Committee
- TC3
- Place
- Cape Town, SOUTH AFRICA
- Time
- 3 February 2014 - 6 February 2014
- Website
- http://conferences.imeko.org/index.php/tc3-5-22/2014