Skip to main content

A NON-IMAGING OPTICAL SYSTEM FOR CHARACTERISATION OF BALL-SHAPED MICRO-INDENTERS

Febo Menelao, Zhi Li, Uwe Brand, Andre Felgner

Abstract

To quantitatively determine the tip radius of spherical and conical indenters, a practical non-contact radius measurement system is presented. Realizationof the optical system has been detailed, including its theoretical and actual resolution and error sources. For the purpose of experimental investigation of the performance of the nonimaging radius measurement system, a sapphire sphere with a nominal radius of 200 µm has been employed. Detai led data analysis indicates that high-order form error of the object under test might be one of the key error sources within the calibration setup.

Keywords
confocal microscopy, micro- and nanoindentation, indenter, Rockwell indenter, ball-shaped indenter, radius measurement
Download
IMEKO-TC5-2014-002.pdf
DOI
-
IMEKO TC
TC5 - Hardness Measurement

Event details

Event
TC5 Conference 2014
Technical Committee
TC5
Place
Cape Town, SOUTH AFRICA
Time
3 February 2014 - 6 February 2014
Website
http://conferences.imeko.org/index.php/tc3-5-22/2014

Back to the proceedings