STUDY OF SHOCK MEASUREMENTS USING HOMODYNE AND HETERODYNE LASER INTERFEROMETERS
Abstract
A comparison measurement using homodyne and commercial heterodyne laser interferometers was investigated in the primary shock facility of NMIJ. Each derived shock waveform was calculated by applying a two stage differentiation process with 4<sup>th</sup> order Butterworth lowpass filter. As results, the time difference and the deviation between peak accelerations for two amplitude levels were respectively less than 1 µs and quite less than 0.01 %.
Event details
- Event
- TC22 Conference 2014
- Technical Committee
- TC22
- Place
- Cape Town, SOUTH AFRICA
- Time
- 3 February 2014 - 6 February 2014
- Website
- http://conferences.imeko.org/index.php/tc3-5-22/2014