ACCURATE LINEAR THERMAL EXPANSION COEFFICIENT DETERMINATION BY INTERFEROMETRY
Abstract
In precision length measurements the accurate and traceable value of the linear thermal expansion coefficient (LTEC) is needed. A device for interferometric determination of the LTEC of gauge blocks has been constructed. Minimum temperature gradients in a gauge block with 500 mm maximum length and relatively fast operation were the objectives of this project. Based on uncertainty analysis LTEC can be measured with a standard uncertainty of 0,02·10<sup>-6</sup> 1/K for 100 mm gauge blocks.
Event details
- Event
- XVII IMEKO World Congress
- Place
- Dubrovnik, CROATIA
- Time
- 22 June 2003 - 28 June 2003
- Website
- http://www.hmd.hr/imeko/