A new system to deliver vibration, measure motion and EMG reaction for forearm muscles

Mario Cesarelli, Giovanni d’Angelo, Maria Romano, Paolo Bifulco, Fabrizio Clemente

Abstract

This study aims to measure motion and neuromuscular activity of the forearm flexor muscle in response to mechanical, vibratory stimuli. In particular, the vibration frequency that maximizes the neuromuscular response was searched. A dedicated measuring system was realised: the mechanical stimulus was provided to subject’s hand by a controlled, vibrating plate; the forearm muscles response was detected through sEMG and tri-axial accelerometric signals. Sinusoidal vibrations at different frequencies in the range 14–32 Hz were delivered. The RMS of the sEMG signal and a accelerometric transfer function (i.e. ratio between acceleration of the plate and acceleration of the target muscle) were computed for the frequency range and then analysed. Preliminary results on six subjects show different mechanical and neuromuscular behaviour between the subjects but also intra-subject variability due to posture and initial state of the muscle. The EMG response to vibration depends on frequency.

Keywords
Skeletal muscles tissue, vibration, emg, forearm
Download
IMEKO-TC4-2014-284.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2014
Technical Committee
TC4
Place
Benevento, ITALY
Time
15 September 2014 - 17 September 2014
Website
http://www.imeko-tc4-2014.org