Skip to main content

Design, stability analysis and uncertainty contribution of a voltage divider designed for a phase meter

Stanislav Maslan, Martin Sira, Vera Nováková Zachovalová

Abstract

Today uncertainties of a phase difference of two sinusoidal signals is commonly about 10 μrad for non coherent measurements and signal amplitudes up to several volts. Increasing voltage range of a phase meter requires special considerations. The construction of a special voltage divider intended for precise phase difference measurement is presented in this paper. The uncertainty contribution of the voltage divider was determined by repeated measurement and an Allan variance calculation. The result showed out the voltage range of the phase meter increased substantially with very small addition to the total uncertainty of the phase difference.

Keywords
phase measurement, measurement uncertainty, voltage dividers, stability analysis
Download
IMEKO-TC4-2014-192.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details