Skip to main content

Calorimetric sensor to measure local calorific dissipations in the human body

Ch. Jesus, Fabiola Socorro, Manuel Rodriguez de Rivera, Hector Jesus Rodriguez de Rivera, Roberto Sarmiento

Abstract

A calorimetric sensor for medical application has been developed in order to measure surface and localized calorific dissipations in the human body. The instrument tries to evaluate the calorific power that is transmitted by conduction, through a thermopile, between the human body and a thermostat at a constant temperature. The detection area of the prototype is 36 cm2 and the sensor resolution is 50 mW. From the equations of heat transport by conduction, we propose a model based on the sensor decomposition in different domains connected to each other. The model is validated experimentally. The sensor operation domain is indicated and some practical applications to the human body are shown.

Keywords
Calorimetric sensor, Direct calorimetry, Isothermal calorimeters, Medical calorimetry, Non-differential calorimeters
Download
IMEKO-TC4-2014-258.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2014
Technical Committee
TC4
Place
Benevento, ITALY
Time
15 September 2014 - 17 September 2014
Website
http://www.imeko-tc4-2014.org

Back to the proceedings