Skip to main content

New generation of cage-type current shunts at CMI

Věra Nováková Zachovalová, Martin Šíra, Pavel Bednář

Abstract

This paper describes a new generation of cage-type ac/dc current shunts from 30 mA up to 10 A developed using a lumped circuit element model. Comparison of the calculated and measured values shows agreement better than 6 %Omega;/%Omega; in the ac-dc difference at frequencies up to 100 kHz.

Keywords
current shunt, current measurement, shunts modelling, electric variables measurement
Download
IMEKO-TC4-2014-309.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
TC4 Symposium 2014
Technical Committee
TC4
Place
Benevento, ITALY
Time
15 September 2014 - 17 September 2014
Website
http://www.imeko-tc4-2014.org

Back to the proceedings