DC Voltage Linearity Measurements and DVM Calibration with Conventional and Programmable Josephson Voltage Standards

Regis Pinheiro Landim, Waldemar Guilherme Kürten Ihlenfeld

Abstract

Traditional calibration of dc voltage high accuracy digital voltmeters (DVMs) using conventional and Programmable Josephson Voltage Standards at INMETRO show improvements between 73.7 % and 98.3 % when compared with a regular calibration technique. Some calibration results as well as DVM linearity and gain drift analysis (up to 10 V range) are presented and discussed in this paper.

Keywords
DVM linearity; DVM calibration; Gain Error Drift; Programmable Josephson Voltage Standards