Skip to main content

DYNAMIC CALIBRATION METHODS FOR FORCE TRANSDUCERS

Yusaku Fujii

Abstract

Three Methods for evaluating the dynamic response of force transducers against varying force are described in this paper. In all methods, the inertial force of a mass is used as the known dynamic force, and this reference force is applied to a force transducer under test. The inertial force is measured highly accurately as the product of the mass and the acceleration. An aerostatic linear bearing is used to obtain linear motion with sufficiently small friction acting on the mass (i.e., the moving part of the bearing). Three experimental setups were built for the dynamic calibration against an impact force, an oscillation force and a step force. Ways of establishing dynamic calibration methods are also discussed.

Keywords
metrology, standards, mechanical quantities, dynamic, force, calibration, sensor, varying force, inertial, impact, oscillation, step, optical interferometer, pneumatic linear bearing
Download
PWC-2006-TC3-015u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity