Skip to main content

A NEW CALIBRATION METHOD WITH STATIC LOADS FOR PIEZOELECTRIC FORCE TRANSDUCERS

Oliver Mack

Abstract

This paper presents a new method for calibrating piezoelectric force transducers with static forces. Load cycles with pilgrim-steps and a special evaluation of the measuring results permit the determination of transducer coefficients that are independent of the linear drift of the charge amplifier. Consequently, the coefficients are independent of the measurement time and the time of load change of the force standard machine in use. It is shown that the new calibration method contributes essentially reducing the uncertainty of piezoelectric force transducers that are used in quasistatic and dynamic precision force measurements.

Keywords
piezoelectric force transducer, calibration method, static force
Download
PWC-2006-TC3-018u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XVIII IMEKO World Congress
Place
Rio de Janeiro, BRAZIL
Time
17 September 2006 - 22 September 2006
Website
http://www.metrologia2006.org.br

Back to the proceedings