UNCERTAINTY ANALYSIS AND LONG-TERM STABILITY INVESTIGATION OF THE GERMAN PRIMARY HIGH-PRESSURE NATURAL GAS TEST FACILITY pigsar
Abstract
The high-pressure gas flow meter test facility pigsar operated by Ruhrgas AG serves for testing and calibration of turbine-wheel and ultrasonic meters for natural gas in the pressure range between 14 and 50 bar and flow rates ranging from 8 to 6500 m³ / h. Pigsar is the national standard of high-pressure natural gas flow under supervision of the German national metrological institute PTB. It represents and disseminates the unified German-Dutch reference value for the unit of volume for high-pressure natural gas. The uncertainty analysis of the test facility presented here was one of the main pre-requisites for the creation of the harmonised reference value.
Event details
- Event
- FLOMEKO 2003
- Technical Committee
- TC9
- Place
- Groningen, NETHERLANDS
- Time
- 12 May 2003 - 14 May 2003
- Website
- www.flomeko2003.nl