Skip to main content

THE MIKES MEASURING SYSTEM FOR GAS MASS FLOW

Sampo Sillanpää, Martti Heinonen

Abstract

Due to the increasing need for traceable gas flow measurements in Finland, the development of a national measurement standard for gas mass flow was initiated at the Centre for Metrology and Accreditation (MIKES) in 2002. A primary calibration system based on dynamic weighing was constructed to provide traceability directly to the national mass and time standards. The MIKES gas flow measuring system includes also a commercial calibrator based on laminar flow elements. It is used as a working standard when calibrating mass flow meters for customers. The calibration range encompasses 0,4 mg / s to 625 mg / s. The relative standard uncertainty of the gravimetric system is between 0,2 % and 0,4 %.

Download
IMEKO-TC9-2003-026.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2003
Technical Committee
TC9
Place
Groningen, NETHERLANDS
Time
12 May 2003 - 14 May 2003
Website
www.flomeko2003.nl