Skip to main content

DEVELOPMENT OF A VERY SMALL ULTRASONIC FLOWMETER FOR LIQUIDS

H. Ishikawa, M. Takamoto, K. Shimizu, H. Monji, G. Matsui

Abstract

In this study, small disk ultrasonic transducers were used to develop a very small ultrasonic flowmeter with a measuring pipe diameter of less than 1 mm. The feasibility of using an ultrasonic flowmeter to measure a liquid flow rate below 1 mL/min was investigated experimentally. Calibration results of the developed ultrasonic flowmeter were almost all scattered within the range predicted from the standard deviation of measurements. The flowmeter can stably measure a flow rate of less than 1 mL/min with a standard deviation of about 0.01 mL/min. This result suggests that the flowmeter can be very useful for semiconductor, medical and chemical industries.

Keywords
ultrasonic flowmeter, very low flow rate, zero stability, cross-correlation, small disk ultrasonic transducer
Download
IMEKO-TC9-2000-095.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2000
Technical Committee
TC9
Place
Salvador, Bahia, BRAZIL
Time
4 June 2000 - 8 June 2000

Back to the proceedings