CALIBRATION AND UNCERTAINTY EVALUATION OF FLICK STANDARD USING COORDINATE-MEASURING MACHINE

Yohan Kondo, Youichi Bitou, Osamu Sato, Makoto Abe

Abstract

Flick standard is a cylindrical artifact with a flat part for calibrating the magnification of a probe of roundness measuring-instruments. A probe magnification of roundness measuring-instruments is evaluated by measuring a calibrated flick value of a flick standard. The flick value is a distance between a cross-section circle and a cross-section line. In factory field, it is required that the flick value is calibrated within 0.1 µm with respect to several micrometer to several hundred micrometer. The flick value is generally calibrated by an accurate roundness measuring-instrument. It has a linear error of a probe with about 0.1% or more gain error. When the flick value is 100 µm, the measurement uncertainty is 0.1 µm or more. We calibrated a flick standard using an ultra-high accurate coordinate measuring instrument (CMM) with a laser interferometer to ensure the measurement uncertainty within 0.1 µm. The length measurement system by the laser interferometer is according to the abbe principle. In this paper, we describe the calibration method of a flick standard using the CMM. We calibrated the flick standard using a multiple-measurement technique to eliminate the systematic error of the CMM and estimated its uncertainty based on the analysis of variance. Finally, the proposed calibration method is validated through experiments.

Keywords
Flick standard, roundness measuring- instrument, coordinate measuring-machine, interferometer, uncertainty
Download
IMEKO-TC14-2014-59.pdf
DOI
-
IMEKO TC
TC14 - Measurement of Geometrical Quantities

Event details

Event
TC14 LMPMI Symposium 2014
Technical Committee
TC14
Place
Tsukuba, JAPAN
Time
3 September 2014 - 5 September 2014
Website
http://www.lmpmi2014.jp/

Back to the proceedings