A CURRENT SOURCE FOR CALIBRATION OF LOW-CURRENT METERS

Luca Callegaro, Vincenzo D’Elia

Abstract

The need for accurate calibration of current meters in the picoampere range is continuously increasing. A traceable current source for calibration of meters in the 100 fA – 100 pA current range is here presented. The source is based on a low-frequency generator ramp generator, which charges and discharges a gas-dielectric capacitor. Current traceability is given by the measurement of voltage parameters with a sampling voltmeter, and by calibration of the capacitor at audio frequency. The source has been employed in the EUROMET.EM-S24 "Comparison of small current sources" supplementary comparison.

Keywords
current measurement, calibration, signal generators
Download
PWC-2006-TC4-003u.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XVIII IMEKO World Congress
Place
Rio de Janeiro, BRAZIL
Time
17 September 2006 - 22 September 2006
Website
http://www.metrologia2006.org.br

Back to the proceedings