A CURRENT SOURCE FOR CALIBRATION OF LOW-CURRENT METERS
Abstract
The need for accurate calibration of current meters in the picoampere range is continuously increasing. A traceable current source for calibration of meters in the 100 fA – 100 pA current range is here presented. The source is based on a low-frequency generator ramp generator, which charges and discharges a gas-dielectric capacitor. Current traceability is given by the measurement of voltage parameters with a sampling voltmeter, and by calibration of the capacitor at audio frequency. The source has been employed in the EUROMET.EM-S24 "Comparison of small current sources" supplementary comparison.
Event details
- Event
- XVIII IMEKO World Congress
- Place
- Rio de Janeiro, BRAZIL
- Time
- 17 September 2006 - 22 September 2006
- Website