PRIMARY SINUSOIDAL CALIBRATION OF FORCE TRANSDUCERS UP TO 2 KILOHERTZ
Abstract
We report the implementation of a swept-sine dynamic force calibration system at NIST. This system is designed to perform primary calibration of force transducers at frequencies up to 2 kHz and forces up to 2 kN. We demonstrate the use of this system to measure the dynamic sensitivity of a strain-gauge based force transducer up to 2 kHz. The current dominant source of uncertainty in the system is angular motion of the calibration assembly.
Event details
- Event
- XXI IMEKO World Congress
- Place
- Prague, CZECH REPUBLIC
- Time
- 30 August 2015 - 4 September 2015
- Website
- http://imeko2015.org/