Skip to main content

THE PERFORMANCE EVALUATION OF LABORATORIES PERFORMING CALIBRATION OF NON-AUTOMATIC WEIGHING INSTRUMENTS (NAWI)

Adriana Vâlcu, Alina Taină, Roberta Todor

Abstract

In 2014, a proficiency testing in the field of NAWI calibration was carried out at National Institute of Metrology (INM) from Romania. The accredited provider of this proficiency testing scheme was ROLAB Association. The subject of the comparison consisted in a weighing instrument having Max 6000 g and <i>d</i> = 0.01 g. The paper presents the results obtained by eleven participating laboratories in this proficiency testing scheme. All results were in agreement with the results of INM.

Keywords
non automatic weighing instrument (NAWI), normalized error, error of indication, uncertainty attributed to error of indication
Download
IMEKO-WC-2015-TC3-087.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings