Skip to main content

MEASURES OF ELECTROMAGNETIC COMPATIBILITY: RFID AND ELECTROMEDICAL DEVICES IN THE HOSPITAL ENVIRONMENT

Marcantonio Catelani, Michele Chini, Andrea Zanobini

Abstract

The objective of this paper is to assess whether some Electromedical devices are susceptible to electromagnetic waves emitted from sources for remote identification. We examine the occurrence of electromagnetic interference with a measuring method developed by the University of Florence from protocol ANSIC63.18. We have also conducted several in-situ tests, to characterize the EM appliances, from the point of view of electromagnetic compatibility, directly in to the hospital environment.

Keywords
Electro-Magnetic Compability (EMC), Radio Frequency IDentification (RFID), Medical device, essential performance, ElectroMagnetic Interference (EMI), in-situ test
Download
IMEKO-WC-2015-TC4-108.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings