Skip to main content

EFFECTS OF STRAY CAPACITANCE TO GROUND IN BIPOLAR WATER IMPEDANCE MEASUREMENTS BASED ON CAPACITIVE ELECTRODES

Carles Aliau-Bonet, Ramon Pallas-Areny

Abstract

Liquid impedance measurements based on capacitive (or contactless) electrodes overcome electrode polarization problems but are affected by stray capacitance from the material being measured to ground, the same as measurements with direct-contact electrodes. This study shows that the effects of that capacitance depend on the impedance being measured and for bipolar impedance measurements they increase when the ratio between that stray capacitance and electrode capacitance increases.

Keywords
capacitive electrodes, stray capacitance to ground, impedance measurements
Download
IMEKO-WC-2015-TC4-138.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings