VERIFICATION RESULTS OF LEEB HARDNESS INDICATING DEVICES FROM DIFFERENT MANUFACTURERS USING THE LEEB HARDNESS REFERENCE VOLTAGE SIGNAL
Abstract
The indicating device of a Leeb hardness tester is an important device that determines the performance of tester. As such, the indicating device needs verification just like the depth-measuring system of a Rockwell hardness tester. This paper presents the verification results of indicating devices from different manufacturers. For the verification, a calibrator to output the reference voltage signal was designed and used for the test. The results showed that the maximum deviation was 24 HLD.