DEVELOPMENT OF VOLUMETRIC MICRO-FLOW CALIBRATION SYSTEM USING FPGA FOR MEDICAL APPLICATION

Bernadus H. Sirenden, Ghufron Zaid, Prawito Prajitno, Hafid

Abstract

To answer the need for traceability in micro flow area for medical industry, Flow Laboratory of Metrology LIPI in cooperation with Physics Department of University of Indonesia, developed a compact standard based on piston displacement concept. The standard consists of glass syringe and linear actuator controlled by Field Programmable Gate Array (FPGA). The volume generation uncertainty is 0.6% to 0.7% at 6.5 mL to 13.4 mL. Flow-rate uncertainty is 1.6% to 11.6% at 90.3 mL/hr to 1177.1 mL/hr.

Keywords
micro-flow, syringe pump, FPGA