Skip to main content

DEVELOPMENT THE PRIMARY VELOCITY STANDARD FOR THE CALIBRATION OF LDA SYSTEM

Wu Jian, Nikolay Semin

Abstract

Air velocity measurements are required to have the measurement traceability to the SI unit of length and time. This can be realised through the setup of a primary velocity standard. NMC has designed and developed the primary velocity standard using the spinning disc for the calibration of an LDA system. The system is capable of measuring the velocity from 0.2 m/s to 40 m/s with the measurement uncertainty of 0.1% to 0.4% of reading.

Keywords
Laser Doppler Anemometry (LDA), air velocity, primary standards, uncertainty evaluation
Download
IMEKO-WC-2015-TC9-229.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings