REALIZATION OF THE ITS-90 FOR RANGE 232 °C – 962 °C ON KE LP4 AT METROLOGY-LIPI
Abstract
The realization of ITS-90 has been done at Metrology-LIPI using the radiation thermometer monochromatic KE LP4 80-59 over a temperature range 232 - 962 ° C and wavelength λ = 1.57 µm. The method used is Sakuma-Hattori method using fixed-point blackbody tin, aluminum and silver. The value of the uncertainty of the KE LP4 80-59 at the temperature range for the level of confidence of 95% and a coverage factor of <i>k</i> = 2, is expressed by the equation: <i>U</i>(<i>t</i>) = 1.067461375 - 0,008000541 · <i>t</i> + 2,586471 · 10<sup>-5</sup> · <i>t</i><sup>2</sup> -3,33928 · 10<sup>-8</sup> · <i>t</i><sup>3</sup> + 1,497270 · 10<sup>-12</sup> · <i>t</i><sup>4</sup>.