Skip to main content

DETERMINATION OF FORM MEASURING MACHINE DISPLACEMENT SENSOR CHARACTERISTICS WITH A USE OF FLICK STANDARD

Marta Wiśniewska, Sabina Żebrowska-Łucyk

Abstract

A new approach towards calibration of FMM displacement sensor with a use of a flick standard is presented. The main drawback of the prevailing procedure is the fact that instrument’s probe could have been calibrated only for a single value, whereas following a novel method removes this limitation. Therefore, FMM probe linearity deviations may be determined within a full range of sensor. In the paper, the core idea and sample results presenting performance of the proposed method are outlined.

Keywords
form measuring machine, calibration, sensor characteristics, flick standard
Download
IMEKO-WC-2015-TC14-320.pdf
DOI
-
IMEKO TC
TC14 - Measurement of Geometrical Quantities

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings