CALCULATION OF EFFECTIVE AREA BASED ON DIMENSIONAL MEASUREMENT FOR 35 mm DIAMETER PISTON-CYLINDER UNIT OF KIM-LIPI PRESSURE STANDARD

R. Rudi Anggoro Samodro, Adindra Vickar Ega, Hafid

Abstract

The calculation of KIM-LIPI 35 mm diameter piston-cylinder effective area, based on the dimensional measurement has been done successfully. This effective area was calculated by using simplified formula and simple numerical analysis. The purpose is to get a smaller uncertainty as the source of traceability chain for pressure in KIM-LIPI. The both result shows good agreement and only differ 0.6 &times; 10<sup>-6</sup> relatively and it is consistence compare with the result based on crossfloat with relatively less than 2 &times; 10<sup>-6</sup>.

Keywords
effective area, pressure balance, dimensional measurement, simplified formula, numerical analysis
Download
IMEKO-WC-2015-TC16-332.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/