DEVELOPMENT OF A MEASUREMENT SYSTEM FOR DYNAMIC CALIBRATION OF PRESSURE SENSORS

Andrej Svete, Metka Štefe, Andraž Maček, Jože Kutin, Ivan Bajsić

Abstract

The aim of this paper is to present the developed measurement system for dynamic calibration of the pressure sensors at different average pressures. The dynamic pressure generator under discussion consists of the pneumatic cylinder, which is divided by the piston into two pressurized gas chambers. The piston-rod assembly is driven by an electrodynamic shaker. Experimental analyses and mathematical modelling were employed to demonstrate its capabilities and limitations.

Keywords
dynamic calibration, pressure pulsations, pneumatic cylinder, natural frequency
Download
IMEKO-WC-2015-TC16-340.pdf
DOI
-
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings