Skip to main content

ACCURACY EVALUATION IN THIAMIN QUANTIFICATION IN FOOD MATRICES

Cristina Flores, M. Graça Dias, Mariana Santos

Abstract

The performance evaluation of an RP-HPLC method to determine thiamine using different food matrices and reference materials, showed the method suitability to its purpose and to be submitted to accreditation. Intermediate precision evaluate through the relative standard deviation (RSD%) was under 8% for all food matrices tested an all the z-scores obtained for the reference materials were < 1.5.The estimated expanded measurement uncertainty was 13%.

Keywords
thiamin, accuracy, foods, HPLC
Download
IMEKO-WC-2015-TC23-447.pdf
DOI
-
IMEKO TC
TC23 - Metrology in Food and Nutrition

Event details

Event
XXI IMEKO World Congress
Place
Prague, CZECH REPUBLIC
Time
30 August 2015 - 4 September 2015
Website
http://imeko2015.org/

Back to the proceedings