Accurate VOC mixtures at ppb level are needed both for the atmospheric monitoring both for the clean room industries. Research is developing a compact diffusion generator of trace mixtures to be installed on-site. To assure the required accuracy, vials with a low and calibrated diffusion rate DR are necessary. In this paper glass vials were realized, the uncertainty budget of DR was evaluated and experimentally validated. A DR accuracy better than 3% at 200 ng/min release rate was obtained.