Skip to main content

EURAMET cg-18 - Calibration of Balances and Scales - Update on International Distribution and Acceptance

Klaus Fritsch

Abstract

This paper presents the state-of-the-art calibration strategy for non-automatic weighing instruments (NAWI), based on the recently revised guideline EURAMET cg-18. Besides of the uncertainty at calibration, the document provides advice for the so-called uncertainty of a weighing result that describes the device performance during day-to-day use. The concept of minimum weight was added during a recent revision. The minimum weight is the smallest sample quantity that can be weighed on the device, still adhering to a predefined relative weighing tolerance requirement.

Keywords
calibration, non-automatic weighing instruments, minimum weight, harmonization, quality
Download
IMEKO-TC3-2017-008.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
tc3-2017

Back to the proceedings