Skip to main content

Dynamic Calibration of Fatigue Test Machines Using a Dynamic Calibration Device

Rafael S. Oliveira, Renato R. Machado, Claudio A. Koch, Luiz C. Cabral de Freitas, Wellington G. Fernandes, Antonio L. Gama

Abstract

This paper describes the practical application of method B of the ISO 4965-1 standard for the dynamic force calibration of a fatigue test machine. The experiment focuses on detailing the main proceedings involved such as the load series, the frequency sweep range and the data acquisition and processing parameters applied to the Dynamic Calibration Devices (DCDs), used with different compliances. The calibration results are shown and a discussion is proposed.

Keywords
ISO 4965-1, dynamic force calibration, dynamic traceability, fatigue test machines
Download
IMEKO-TC3-2017-024.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
tc3-2017

Back to the proceedings