Improvement of the Primary Low-Frequency Accelerometer Calibration System at INMETRO

G. P. Ripper, C. D. Ferreira, R. S. Dias, G. B. Micheli

Abstract

This paper presents the results of recent improvements of the primary low-frequency accelerometer calibration system at INMETRO. The interferometric system was upgraded, a new software developed and the sine- approximation method was implemented. This new system allows the calibration of the complex sensitivity of accelerometers from 0,2 Hz to 100 Hz with reduced uncertainty.

Keywords
vibration, accelerometer, calibration, interferometry, metrology