Pico-Meter Metrology for the Primary Vibration Calibration
Abstract
This paper introduces the use of a homodyne differential plane mirror interferometer (HDPMI) for development of the primary vibration calibration system. The zero-drift test results showed that it enables displacement measurement in a scale of 10 pico-meters could be feasible. The HDPMI model was shown to be quite successful for the primary vibration calibration even in the frequency range of 5 kHz to 20 kHz. The standard uncertainty of the complex sensitivities from 5 kHz to 20 kHz was to be less than 0.46 % for their moduli and 0.57° for the phase shifts.
Event details
- Event
- TC22 Conference 2017 - Measurement facing new challenges!
- Technical Committee
- TC22
- Place
- Helsinki, FINLAND
- Time
- 30 May 2017 - 1 June 2017
- Website
- http://conferences.imeko.org/index.php/tc3-5-22_2017/2017