Skip to main content

Measurement of Time Delay Caused by Mixer And Lowpass Filter in Heterodyne Interferometer

Ming Yang, Chenguang Cai, Ying Wang, Zhihua Liu, Haijiang Zhu

Abstract

Phase response is a key performance parameter of vibration transducers, which is usually measured by the primary calibration. The typical primary calibration method used analog mixer and lowpass filter to modulate heterodyne laser interferometer signal. There is some time delay caused by the analog devices, which will introduce some phase delay for the measurement of phase response of transducers. This paper proposed a novel time delay measurement method of the analog devices based on high definition oscilloscope (HDO) and phase unwrapping sine approximation method (PUSAM).

Keywords
phase response, primary calibration, phase delay, oscilloscope, sine approximation method
Download
IMEKO-TC22-2017-012.pdf
DOI
-
IMEKO TC
TC22 - Vibration Measurement

Event details

Event
TC22 Conference 2017 - Measurement facing new challenges!
Technical Committee
TC22
Place
Helsinki, FINLAND
Time
30 May 2017 - 1 June 2017
Website
http://conferences.imeko.org/index.php/tc3-5-22_2017/2017

Back to the proceedings