Skip to main content

DENSITY STANDARD LIQUIDS OF NEW CONCEPT BY THE MAGNETIC LEVITATION

Yohei Kayukawa, Ken-ichi Fujii, Yuya Kano, Haruki Sato

Abstract

The major uncertainty factor of the magnetic levitation densimeter, which is one of the most accurate densimeter, is the force transmission error caused by the magnetism of the fluid. We have examined this effect by FEM, and pro-
posed a new method for the magnetic levitation densimeter.

Keywords
densimeter, density standard liquid, magnetic coil levitation
Download
PWC-2006-TC3-036u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
XVIII IMEKO World Congress
Place
Rio de Janeiro, BRAZIL
Time
17 September 2006 - 22 September 2006
Website
http://www.metrologia2006.org.br