Study on Charge Standard for calibrating PD Measurement Equipments

Na Liu, Shengyou Gao, Caijun Yang, Lu Huang

Abstract

The characteristics of a portable charge standard for calibrating partial discharge(PD) measurement equipments are discussed and investigated in this paper. A portable PD calibrator and measuring system was set up by NIM, China. This calibrator and two typical commercial ones are measured by NIM’s digital system. The discharge impulses and the charge generated by the three calibrators are recorded and compared. The comparison results show that the portable PD calibrator is of good characteristic both in the range of charge and in the shape of its impulses and also in that of the repeatability.

Keywords
charge, partial discharge
Download
IMEKO-TC4-2019-018.pdf
DOI
-
IMEKO TC
TC4 - Measurement of Electrical Quantities

Event details

Event
tc4-2019

Back to the proceedings