Skip to main content

DEVELOPMENT OF METAL-CARBON EUTECTIC CELLS FOR CONTACT THERMOMETRY AT KRISS

Yong-Gyoo Kim, Inseok Yang

Abstract

We report the status of development of metalcarbon eutectic cells for contact thermometry at KRISS. Up to now, Co-C, Ni-C, Fe-C and Pt-C eutectic cells have been fabricated and their performance has been tested. The uncertainty of Co-C melting point using a Type B thermocouple was 0.9°C (<i>k</i> = 2). The melting plateau of the Fe-C eutectic cell has shown a dependence on the prefreezing set temperature.

Keywords
metal-carbon eutectic, thermocouple thermometry, fixed points
Download
PWC-2006-TC12-006u.pdf
DOI
-
IMEKO TC
TC12 - Temperature and Thermal Measurements

Event details

Event
XVIII IMEKO World Congress
Place
Rio de Janeiro, BRAZIL
Time
17 September 2006 - 22 September 2006
Website
http://www.metrologia2006.org.br

Back to the proceedings