Skip to main content

Intercomparison tests of gas flow by Bell Prover

Wang Chi, Li Fang, Meng Tao, Xu Yinghua, Yang Youtao, Jiang Ning, Wu Weilong, Sang Xiaoming, Chen Mei, Zhou Bingzhi, Zhang Jinming, Shen Wenxin, Xu Ling

Abstract

This paper describes the intercomparison tests carried out by 10 labs around China from 2003 to 2004. We compared the bell provers with a flow range from 0. 6 to 6 m³/h by using 2 roots flow meters. Test result and diagram are showed in this paper. We analyzed the influence of temperature control and the seal liquid of bell provers on test.

Keywords
Intercomparison, Bell Prover, Temperature, Seal liquid
Download
IMEKO-TC9-2004-022.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
tc9-2004

Back to the proceedings