Digital post-distortion of an ADC analog front-end for gamma spectroscopy measurements
Abstract
This work presents the experimental characterization and digital post-distortion (i.e., digital linearization) of a MHz-range ADC analog front-end for gamma radiation spectroscopy measurements. The front-end was firstly characterized by means of a conventional approach yielding Total Harmonic Distortion (THD) and Effective Number Of Bits (ENOB). Then, it was tested using μs-range pulsed excitations, which are the typical waveforms encountered in gamma spectroscopy. After applying digital post-distortion, the linearized receiver shows a THD reduced by more than 20 dB, a substantial improvement in ENOB, and a higher accuracy in the acquisition of pulsed waveforms.
Event details
- Event
- tc4-2020