Mohamad Syahadi, Hayati Amalia, Agah Faisal

Abstract

A method for calibrating a solid state DC voltage standard in Laboratory of National Measurement Standards for Electricity and Time (NMS ET Lab) of National Standardization Agency of Indonesia is presented in this paper. The measurement and its evaluation are carried out based on differential measurement method to the test of solid state DC voltage standard by implementing calibration values from Bureau International des Poids et Mesures (BIPM) at the nominals of 1.018 V and 10 V. Corrections due to temperature, pressure, humidity coeficient and drift are also examined. Based on evaluation result, actual values for the 1.018 V and 10 V of the test were found to be 1.0179897 V and 9.999857 V with their respective measurement uncertainties of 1 μV and 17 μV.

Keywords
Actual value, Calibration, Differential measurement method, Solid state DC voltage standard, Measurement uncertainty