Skip to main content

METROLOGICAL RELIABILITY OF HIGH FREQUENCY SURGICAL EQUIPMENT

Rosário P A Palomares, Paula M P Gouvêa, Elisabeth Costa Monteiro

Abstract

The metrological reliability of high frequency surgical equipment in use at three hospitals in Rio de Janeiro, Brazil was evaluated. The units tested did not satisfy safety and performance requirements, indicating that (post-sale) periodic verifications are essential, and that crest factor analysis should be included in technical standards.

Keywords
legal metrology, metrological reliability, medical electrical equipment, high frequency surgical equipment, electrosurgical unit
Download
PWC-2006-TC13-018u.pdf
DOI
-
IMEKO TC
TC13 - Measurements in Biology and Medicine

Event details

Event
XVIII IMEKO World Congress
Place
Rio de Janeiro, BRAZIL
Time
17 September 2006 - 22 September 2006
Website
http://www.metrologia2006.org.br

Back to the proceedings