A preliminary study on the blind angle estimation for Quality Assessment of Color Doppler ultrasound diagnostic systems
Abstract
Quality Controls (QCs) for the evaluation of Color Doppler (CD) performance play a pivotal role since CD is one of the most widely used Doppler techniques in medical imaging. Among the recommended test parameters, directional accuracy at 90° is included, even if its assessment is carried out visually. Therefore, in this study, a novel CD test parameter, called blind angle, has been proposed and defined as it allows testing the Doppler frequency shift dependency on the insonation angle. Moreover, a novel automatic method for the estimation of this parameter through the post-processing of CD videos has been proposed. Data have been collected from a single ultrasound diagnostic system equipped with a phased array probe. Tests have been carried out for two probe frequencies and three constant flow rate regimes set on a flow phantom. Based on the preliminary promising outcomes, further studies are going to be carried out.
Event details
- Event
- TC4 Symposium 2022
- Technical Committee
- TC4
- info@imeko-tc4-2022.org
- Place
- Brescia, ITALY
- Time
- 12 September 2022 - 14 September 2022
- Website
- http://www.imeko-tc4-2022.org/
- Proceedings
- https://www.imeko.org/publications/tc4-2022