ESTABLISHMENT OF THAILAND PRESSURE STANDARD FROM 1.5 kPa TO 500 Mpa
Abstract
This paper describes the method used to characterize the zero-pressure effective area (<i>A</i><sub>0</sub>) and pressure distortion coefficient (<i>λ</i>) of 200MPa-CCPG oil piston-cylinder unit (PCU) in order to realize the national pressure scale from 1.5 kPa to 500 MPa. The <i>A</i><sub>0</sub> was determined by cross-floating method from a set of primary gas pressure standard, while the H-W method was used for calculating <i>λ</i>. A set of three oil PCUs was maintained as a primary pressure standard and used for scaling up and down pressure to determine <i>λ</i> for the other PCUs. The relative standard uncertainty of 200MPa-CCPG pressure standard is 3.0 · 10-5.
Event details
- Event
- TC16 Conference 2022
- Technical Committee
- TC16
- imeko2022@fsb.hr
- Place
- Cavtat-Dubrovnik, CROATIA
- Time
- 11 October 2022 - 13 October 2022
- Website
- https://conferences.imeko.org/event/1/