Skip to main content

INTERNAL COMPARISON BETWEEN A REFERENCE TYPE TORQUE CALIBRATION MACHINE AND A 10 N·m DEADWEIGHT TORQUE STANDARD MACHINE

A. Nishino, M. Hamaji, K. Ogushi

Abstract

A reference torque calibration machine (RTCM) has been developed to validate the calibration of efficient torque transducers. In this study, the calibration and measurement capability (CMC) of the RTCM was evaluated based on the technical guideline. We also compared the RTCM and a 10 N·m deadweight torque standard machine at the National Metrology Institute of Japan (NMIJ) using a torque measuring device with a nominal capacity of 1 N·m.

Keywords
torque; calibration; reference type torque calibration machine; internal comparison; uncertainty
Download
IMEKO-TC3-2022-010.pdf
DOI
10.21014/tc3-2022.010
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
TC3 Conference 2022
Technical Committee
TC3
Email
imeko2022@fsb.hr
Place
Cavtat-Dubrovnik, CROATIA
Time
11 October 2022 - 13 October 2022
Website
https://conferences.imeko.org/event/1/

Back to the proceedings