Skip to main content

COMPARISON OF TWO 1 kN FORCE TRANSDUCERS AT CONTINUOUS CALIBRATION

M. Hiti

Abstract

This paper describes a comparison of simultaneous continuous force introduction to two strain-gauge force transducer in a force calibration machine. The continuous force was generated by a single loading/unloading step of the full load of the 1 kN deadweight force calibration machine. Loading rates from 0.3 kN/s to 3.8 kN/s were applied. The results are presented for different loading speeds and for different amplifier types (225 Hz and 4.8 kHz carrier frequency).

Keywords
calibration; continuous; force; error
Download
IMEKO-TC3-2022-053.pdf
DOI
10.21014/tc3-2022.053
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
TC3 Conference 2022
Technical Committee
TC3
Email
imeko2022@fsb.hr
Place
Cavtat-Dubrovnik, CROATIA
Time
11 October 2022 - 13 October 2022
Website
https://conferences.imeko.org/event/1/

Back to the proceedings